Welcome to Shenzhen Meiruike Electronic Technology Co., Ltd.
Message | About us | Chinese Tel:400-876-9388
CN
美瑞克电子

Category

X

RK9920AY Programmable Withstand Voltage Tester

X
  • Product introduction
     
    This series of program-controlled withstand voltage tester adopts high-performance safety tester designed by high-speed MCU and large-scale digital circuit. The output voltage, rise and fall of output voltage and frequency of output voltage are completely controlled by MCU. It can display breakdown current value and voltage value in real time, and has software calibration function. It is equipped with PLC interface, RS232C, RS485, USB device and USB Host interface can be used to form a comprehensive test system with computer or PLC. Rk99 series of program-controlled medical withstand voltage tester has all the functions of the conventional voltage withstand tester, and the arc (flashover) detection function is added. Through the external oscilloscope, the "flashover" phenomenon of the electrical equipment under test can be detected intuitively, accurately, quickly and reliably. When there is no "flashover" phenomenon in the tested electrical equipment, the oscilloscope displays a stable "Lissajous figure, that is, a closed circle In case of "flashover" of the tested electrical equipment, a larger "flashover" phenomenon of electrical equipment will appear in the edge green of Li Shayu's figure to ensure the safety of the tested electrical equipment. This product is also suitable for testing various electrical equipment with flashover defects. It is an indispensable withstand voltage tester for medical electrical equipment manufacturers, maintenance departments, users, product quality inspection departments and technical supervision departments.
     
    This instrument conforms to GB9706.1-2020 (ec60601-1:2012) medical standard, safety of household and similar electrical appliances Part 1: General requirements information technology equipment ul60065, gb8898 / IEC60065 audio, video and similar electronic and mechanical safety requirements - Safety requirements for electrical equipment for measurement, control and experiment - Part 1: General requirements.
     
    Application area
     
    1. Medical equipment: all kinds of new medical equipment and medical supporting equipment, heart monitoring, medical imaging, biochemical analysis instrument, blood pressure meter and thermometer and other home medical equipment
    2. Diagnosis and treatment equipment: X-ray diagnostic equipment, ultrasonic diagnosis, nuclear medicine, endoscope system, five sense organs treatment equipment, dynamic analysis treatment equipment, low-temperature refrigeration equipment, dialysis treatment equipment, first-aid equipment
    3. Ward nursing equipment and appliances: all kinds of beds, cabinets, operating chairs, beds, etc
    4. Auxiliary equipment: medical nursing data and image processing equipment, rehabilitation equipment, special equipment for the disabled, etc
    5. Oral medical equipment: dental diagnostic medical equipment, dental surgical equipment, dental technician equipment, medical magnetic resonance equipment
     
    Performance characteristics
     
    1. The AC / DC withstand voltage function adopts DDS digital signal synthesis technology to produce accurate, stable, pure and low distortion waveform
    2. Adjustable high voltage rise and fall time to meet the requirements of different test objects, with arc detection function, test results can be saved synchronously
    3. It has dual frequency comprehensive test, frequency range of 50 Hz, 60 Hz humanized operation interface, support digital key input, dial input, operation more concise
    4. Complete operation help prompt, can effectively improve user efficiency, support character type file name input, the maximum length of file name is 12 characters
    5. The test steps and system status information are displayed synchronously, which is convenient to understand the detailed information of the test steps and the system status
    6. Chinese and English bilingual operation interface, to meet the needs of different users, support large capacity storage, adapt to different test application requirements, with oscilloscope interface
    7. It can monitor the phenomenon of arc and flashover
    8. Gb9701. - 2020 (ec60601-1:2012) test voltage requirements: no flashover or breakdown shall occur during the test.
    9. The oscilloscope interface is specially added to the medical pressure tester to observe the flashover and arcing phenomenon through Lissajous figure

  • Parameter / model Program-controlled insulation withstand voltage tester Programmable withstand voltage tester
    RK9920 RK9910 RK9920AY RK9910AY RK9920BY RK9910BY
    Withstand voltage test
     
    Output voltage (KV) AC:0.05-5.00 DC:0.05-6.00 AC:0.05-5.00 DC:0.05-6.00 AC:0.05-5.00
    Test accuracy ± (2.0% setting + 2V)
    Output accuracy ± (2.0% setting + 5V) No load
    Test current
    (MA)
    AC:0.001mA-20mA
    DC:0.1uA-10mA
    AC:0.001mA-10mA
    DC:0.1uA-5mA
     
    AC:0.001mA-20mA
    DC:0.1uA-10mA
    AC:0.001mA-10mA
    DC:0.1uA-5mA
    AC:0.001mA-20mA
     
    AC:0.001mA-10mA
    Test accuracy ± (2.0% of reading + 5 words)
    Insulation test
     
    The output voltage
    (KV)
    DC:0.050-1.00
     
    / / / /
    Test accuracy ±(2.0% setting + 2V) / / / /
    Test resistance  
    0.1MΩ-10GΩ
    0.2MΩ-10GΩ / / / /
    Test accuracy 1.0MΩ-10GΩ
    ±20%
    0.1MΩ-999MΩ ±10% / / / /
    Discharge function Automatic discharge after test / / / /
    Arc detection          
    Measuring range Corresponding current 1mA-20mA
    testing time 0.1S-999.9S
    Output frequency 50Hz/60Hz
    Input characteristics 115V/230V±10%  50Hz/60Hz
    Test  alarm Buzzer, LCD, FAIL indicator
    Keyboard lock Independent keyboard lock key
    screen size 5 inch TFT LCD
    Communication Interface HANDLER, RS232, RS485, USBDEV (computer interface), USBHOST (U disk interface)
    Voltage rise time 0.1S-999.9S
    Test time setting
    (AC / DC)
    0.2S-999.9S
    Voltage drop time 0.1S-999.9S
    waiting time 0.2S-999.9S
    Memory 16M  Flash, 50 test steps per file
    size(W×H×D 350mm×110mm×430mm 
    weight 15kg 14kg 15kg 14kg 15kg 14kg
    Accessories Test lead, ground lead, power lead,Cut connector DB9
    Optional accessories
    16g U disk (including PC software), RS232 to USB cable, USB to square port cable

  • Model picture type summary
    RK8N+
    Standard configuration
     
    The instrument is equipped with uncontrolled high pressure rod as standard, which can be purchased separately.
    RK00002
     
    Standard configuration
     
    The instrument is equipped with test line as standard, which can be purchased separately.
    RK00003
    Standard configuration
     
    The instrument is equipped with test line as standard, which can be purchased separately.
    RK00004
    Standard configuration
     
    BNC line is provided as standard and can be purchased separately.
    RK20
    Standard configuration
     
    The instrument is equipped with DB9 as standard, which can be purchased separately.
     
    RK00001
    Standard configuration
     
    The instrument is equipped with American standard power cord, which can be purchased separately.
     
    Certificate and warranty card
     
    Standard configuration
     
    The instrument is equipped with standard certificate and warranty card.
     
    Factory calibration certificate
     
    Standard configuration
     
    Calibration certificate of standard equipment.
     
    instructions
     
    Standard configuration
     
    The instrument is equipped with standard product instructions.
     
    PC software Optional The instrument is equipped with 16g U disk (including upper computer software).
    RS232 to USB cable Optional The instrument is equipped with RS232 to USB cable (upper computer).
    USB to square port cable Optional The instrument is equipped with USB square port connecting cable (upper computer).

  • RK9920AY Programmable Withstand Voltage Tester Click to download the instruction manual
    /uploads/soft/201021/RK99.doc
  • Copyright © 2020 Shenzhen Meiruike Electronic Technology Co., Ltd., All Rights Reserved 粤ICP备07050961号-1 网站建设:深圳北易